Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–299
TRIGGER:MAIN:LOGIC:SETHOLD:DATA:SOURCE CH2
selects Channel 2 as the source for the main logic trigger set/hold.
TRIGger:MAIn:LOGIc:SETHold:HOLDTime
TDS 500C, 600B & 700C only
Sets or queries the main logic trigger hold time. This is equivalent to selecting
Setup/Hold Times from the main Trigger menu and Hold Time in the resulting
side menu.
Trigger
TRIGger:MAIn:LOGIc:SETHold:HOLDTime <NR3>
TRIGger:MAIn:LOGIc:SETHold:HOLDTime?
TRIGger
:
:
LOGIc
MAIn
:
:
?
<Space>
<NR3>
HOLDTime
SETHold
<NR3> specifies the hold time setting in seconds. Positive values for hold time
occur after the clock edge. Negative values occur before the clock edge.
TRIGGER:MAIN:LOGIC:SETHOLD:HOLDTIME 200 E-12
sets the main logic trigger sethold holdtime to 200 nanoseconds.
TRIGger:MAIn:LOGIc:SETHold:SETTime
TDS 500C, 600B & 700C only
Sets or queries the main logic trigger set time. This is equivalent to selecting
Setup/Hold Times from the main Trigger menu and Setup Time in the resulting
side menu.
Trigger
TRIGger:MAIn:LOGIc:SETHold:SETTime <NR3>
TRIGger:MAIn:LOGIc:SETHold:SETTime?