Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–77
CH<x>:PROBEFunc:EXTDBatten:<NR3>
PROBEFunc
CH
:<x>
EXTDBatten
:
?
<Space>
<NR3>
<NR3> is the attenuation value specified in the range –120.00 dB to 120.00 dB.
The default is 0.00 dB.
CH2:PROBEFUNC:EXTDB 2.5
sets the external attenuation to 2.5 dB.
CH<x>:SCAle
Sets or queries the vertical gain of the specified channel. This is equivalent to
setting Fine Scale in the Vertical menu or adjusting the front-panel Vertical
SCALE knob.
Vertical
CH1:VOLts
CH<x>:SCAle <NR3>
CH<x>:SCAle?
CH
<x> :
SCAle
?
<Space>
<NR3>
<NR3> is the gain, in volts per division. The range is 100 mV per division to
1 mV per division when using a 1x probe.
CH4:SCALE 100E-03
sets the channel 4 gain to 100 mV per division.
CH2:SCALE?
might return 1.00E+0, indicating that the current V per division setting of
channel 2 is 1 V per division.
Syntax
Arguments
Examples
Group
Related Commands
Syntax
Arguments
Examples