Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual
53
10. Turn the CURSOR 2 knob to place the second cursor at the point
where the waveform c rosses the second graticule line above
center screen. T his is the 90% level of the waveform.
11. The Delta readout in the Cursor Menu is the rise time of the
waveform.
5 divisions
NOTE. The Rise Time measurement is available as an automati c
measurement in the Measure Me nu, described on page 94.
The Rise Time measurement also displays when you select the Rising
Edge option in the A UTOSET Menu. Refer to page 82.