Appendix D: Probe Basics
TDS3000B Serie s User Manual
D-11
Supported Active Probes and Adapters
The oscilloscope suppli es power to acti ve probe s. You ca n use the
following a ctive probes as long as the total power required by the
probes does not exceed the capacity of the oscilloscope. To
determine the tota l probe load, add the load factors for all probes you
want to use. The oscilloscope can supply power to this combination
if the sum of the l oad factors equals 10 or less. All passive probes
have a loa d factor of ze ro.
Active probe Recommended usage Load factor
P6205 FET probe, 750 MHz, 10X 0
P6243 SMT probe, 1 GHz, 10X 0
P5205 High-voltage differential probe,
1300 V, 100 MHz, 50X or 500X
6
P5210 High-voltage differential probe,
5600 V, 50 MHz, 100X or 1000X
6
ADA400A
Differential preamplifier, 10V sensitiv-
ity, DC to 10 kHz
5
AFTDS
Telecom differential 50 adapter
0
AMT75
Telecom 75 adapter
0
TCP202 Current probe, 15 A, DC to 50 MHz 4
013-0278-00 Video display clamp 5
CAUTION. To avoid a measurement error, do not connect act ive
probes with a combined load factor that is greater than 10. The
signal distortion caused by such an overload can be subtle (reduced
gain, dynamic range, or slew rate).

Table of Contents

Other manuals for Tektronix TDS3034B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3034B and is the answer not in the manual?

Tektronix TDS3034B Specifications

General IconGeneral
BrandTektronix
ModelTDS3034B
CategoryTest Equipment
LanguageEnglish

Related product manuals