Telemark Cryogenics Microprocessor Based Cryochillers
Revision 1.0.7 October 2018
Maximum Load (Watts) 2400 3000 3,600
Typical
Pumping Speed (l/sec)
135,000 165,000 200,000
Ultimate Vacuum 2 x10
mbar (2 x10
Pa) 2 x10
mbar (2 x10
Pa) 2 x10
mbar (2 x10
Pa)
Weight (Kg) 243 (485 lbs) 384 (845 lbs) 412 (930 lbs)
Size inch (mm)
L
W
H
24 (559)
22 (610)
59.5 (1511)
35.3 (897)
23.6 (897)
70.25 (1784)
35.3 (897)
23.6 (897)
70.25 (1784)
Power supply
380-444VAC 3 ph 50Hz
460VAC 3 ph 60 Hz
200-230VAC 3 Ph 50/60 Hz
380-444VAC 3 ph 50Hz
460VAC 3 ph 60 Hz
200-230VAC 3 Ph 50/60 Hz
380-444VAC 3 ph 50Hz
460VAC 3 ph 60 Hz
200-230VAC 3 Ph 50/60 Hz
Wire/Cable Connection 3 Phase + Ground (Earth) 3 Phase + Ground (Earth) 3 Phase + Ground (Earth)
Full Load Current draw @
60 Hz (Amps)
@200-230VAC
@380-440VAC
40
20
50
25
60
30
Start Up Max
Current Draw @ 60 Hz
(Amps)
@200-230VAC
@380-440
60
30
60
30
85
45
Compressor Horsepower
(HP)
10 (7.46kW) 13 (9.69kW) 15 (11.19kW)
Water Requirements (l/min)
@ 15°C
@ 25°C
@ 32°C
6
12
30
6
12
30
8
16
30
Water Connections ¾” NPT Female
Water Resistivity All Models >0.1M Chloride Free
Water (Ph) All Models 6.5 - 8
Refrigeration Connections ½” Metal Seal ½” Metal Seal ½” Metal Seal
Max Coil Length (Meters)
Using ½” Cu tube
35 (114.8 ft) 40 (131.2 ft) 50 (164.0 ft)
Max Surface area (M
) 1.39 (≈15 ft
) 1.59 (≈17.1 ft
) 1.99 (≈21.42 ft
)
*Avg. Cool Time (mins) 2-5 2-5 2-5
*Avg. Defrost Time (mins) 2-4 2-4 2-4
∆T Coil in/Coil Out (°C) ≈5 ≈5 ≈5
Digital Remote
Connections
Input Voltage range
18-24VDC (30VDC MAX) 18-24VDC (30VDC MAX) 18-24VDC (30VDC MAX)
Digital Remote
Connections
Out Voltage range
Relay contact closure 200
VAC or 100VDC Max at
300mA
Relay contact closure 200
VAC or 100VDC Max at
300mA
Relay contact closure 200
VAC or 100VDC Max at
300mA
Balance Pressure Range
as shipped from Telemark
USA facility
(PSI) @20°C
(hand valves closed)
190-220
(1.31-1.52MPa)
250-280
(1.72-1.93MPa)
250-280
(1.72-1.93MPa)
IMPORTANT
.
The above performance figures are based for a unit
running at 60Hz mains power frequency. For units running on 50 Hz
mains frequency the figures above need to be de-rated by 0.83.
Table 2 Model 2400, 3000, 3600 Specifications