EasyManua.ls Logo

TESTO 350 S - Function Key Assignment

TESTO 350 S
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Setting up Control Unit and analyser box measurement system
33
5.5 Assigning function keys
5.5.1 Function bar
·
Measurement menu is activated.
As required:
Use or to change display view so that the function button which is to be
assigned is visible.
1
Press , release and immediately press the function button to be assigned:
-
A selection list of the possible functions appears.
2
Select function and confirm with .
Reversing the assignment:
1
Press menu key , release menu key and then immediately press the defined
function key.
2
Select empty field with and confirm with .
Function bar Control Unit analyser box
Free, non-assigned function button X
Zoom readings XX
Hold current reading X
Display max. values, since switch-on X
Display min. values, since switch-on X
Mean calculation X
Activate volume flow meas. (with a velocity or external X
differential pressure probe or internal pressure probe)
Activate/deactivate velocity measurement (with an external differential X
pressure probe or the internal pressure probe)
Measurement range 40 hPa for internal pressure sensor X
Measurement range 200 hPa for internal pressure sensor X
Zero pressure probe at user-defined probe socket (at least one X
differential pressure probe)
Zero the CO probe X
Start/stop measurement program X X
Determine system configuration X
Save readings XX
Print readings XX
Printer line feed XX
Turbulence calculation (with connected turbulence probe) X
Starts the measuring gas pump and indicates the readings on the display. When X
the function key P Start is clicked on, the function key changes to P Stop
The measuring gas pump stops, the readings are frozen at Hold X
Switches on and zeros velocity rate measurement with Pitot tube and pressure probe X
Manual storing of the current values under the displayed location names. X
Use of the two temperature inputs of the analyser box as separate X
2-channel temperature measurement with differential temperature display.
Activates the separate differential pressure measurement in the analyser box X
(draught measurement).The pump is stopped automatically once “d P” is activated.
Initiates the purging and zeroing phase (1 minute). The device draws fresh air X
through the measurement gas inlet or the fresh air valve (if fitted).
Manual change from measured gas to atmospheric air X
Manual deactivation of an activated CO sensor X
and purging with fresh air.
Manual activation of a deactivated CO sensor X
in the gas path.
Switching the dilution stage (digit in front of x corresponds to the set X
dilution factor)
Switching the HC module on/off X
X
Direct display of all error messages X
V OOn
Save
Delta TT
d PP
Null
COoff
COein
Gas ((air)
1 xx
HC OON
HC OOFF
DIAG
Max
Min
Mean
Vol
m/s
dP2
Pext=0
ppm=0
Start
Stop
Search
Save
Print
LF DDr
dP1
Turb
Zoom
Hold
P SStart
P SStop

Related product manuals