EasyManua.ls Logo

Texas Instruments ADS1298ECG-FE - ADS1 x98 ECG-FE Operation

Texas Instruments ADS1298ECG-FE
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.ti.com
Quick Start
7
SBAU171DMay 2010Revised January 2016
Submit Documentation Feedback
Copyright © 2010–2016, Texas Instruments Incorporated
ADS1298ECG-FE/ADS1198ECG-FE
Table 1. ADS1x98ECG-FE Default Jumper/Switch Configuration
Jumper Default Position Description
JP1 Installed RLD feedback
JP2 Installed 1-2 AVDD selected for bipolar supply operation selected (AVDD = +2.5V)
JP3 Header Not Installed External Vref buffer not connected
JP4 Installed EVM +5V provided from J4 (power header)
JP5 Open PWDN pin controlled from J5 header (pulled up to DVDD)
JP6 to JP14
Header Not Installed (Pins 1-2
shorted on PCB)
DC-coupled input signals
JP15 Installed 2-3 Shield drive is open
JP16 Installed Wilson Central Terminal (WCT) connected to INM for CH1 and CH4-8
JP17 Header Not Installed ECG shield drive
JP18 Installed 2-3 CLK connected to OSC1
JP19 Installed 1-2 OSC1 enabled
JP20 Installed 2-3 AVSS selected for bipolar supply operation (AVSS = -2.5V)
JP21 Installed 1-2 CS connected to DSP via J3.1
JP22 Installed 2-3 START comes from J3.14
JP23 Installed 1-2 CLKSEL set to 0 (ADS1198 uses Ext Master Clock (OSC1))
JP24 Installed 2-3 DVDD supply = 3.3V
JP25 Header Not Installed No external reference selected
JP26
Installed 1-2 (top) WCT connected to CH8- input
Installed 3-4 (bottom) ECG_V1 connected to CH8+ input
JP27
Installed 1-2 (top) WCT connected to CH7- input
Installed 3-4 (bottom) ECG_V5 connected to CH7+ input
JP28
Installed 1-2 (top) WCT connected to CH6- input
Installed 3-4 (bottom) ECG_V3 connected to CH5+ input
JP29
Installed 1-2 (top) WCT connected to CH5- input
Installed 3-4 (bottom) ECG_V4 connected to CH6+ input
JP30
Installed 1-2 (top) WCT connected to CH4- input
Installed 3-4 (bottom) ECG_V2 connected to CH4+ input
JP31
Installed 1-2 (top) ECG_RA connected to CH3- input
Installed 3-4 (bottom) ECG_LL connected to CH3+ input
JP32
Installed 1-2 (top) ECG_RA connected to CH2- input
Installed 3-4 (bottom) ECG_ LA connected to CH2+ input
JP33
Installed 1-2 (top) WCT connected to CH1- input
Installed 3-4 (bottom) ECG_V6 connected to CH1+ input
2.2 ADS1x98ECG-FE Operation
To prepare to evaluate the ADS1298 with the ADS1298ECG-FE, complete the following steps:
1. Verify the jumpers on the ADS1298ECG-FE are as shown in Figure 2 (note that these settings are the
factory-configured settings for the board).
2. Verify that the jumpers on the MMB0 motherboard are configured as shown below:
MMB0 J13A Open
MMB0 J13B Open
Refer to Section B.2 for details about the MMB0 power supply.
3. Install the ADS1298ECG-FE software using the latest software version. The latest software can be
downloaded from the ADS1298ECG-FE product page or ADS1198ECG-FE product page. Double click
the installer and follow the instruction to complete the software installation. For detailed installation
information and screenshots, see Appendix C.

Table of Contents