End-Of-Service (EOS) Mode
www.ti.com
24
SLUUBH1CAugust 2016Revised March 2018
Submit Documentation Feedback
Copyright © 2016–2018, Texas Instruments Incorporated
Battery Gauging
The Smoothing Margin can be set to control the rate of smoothing, but it is recommended to not change
it.
CLASS SUBCLASS NAME TYPE SIZE
MIN
VALUE
MAX
VALUE
DEFAULT
VALUE
UNIT
Gas Gauging Design
Cell Terminate
Voltage
Integer 2 0 5000 2900 mV
EOS Values
EOS SOH Smooth
Start Voltage
Integer 2 0 5000 2800 mV
EOS Data Values
EOS SOH
Smoothing Margin
Integer 1 0 255 128