47
(4) Press the
ESC
/
DISP
key to quit the setting mode and return to the
normal condition. Ab is displayed on the 7-segment MENU LED,
indicating that the tracking function is activated.
Display when tracking function is activated
5-5-10. Master-slave function (10)
CAUTION
The Hi-Ωfunction (03) is not available for the master-slave operation.
Which is only available on the same models in series connection.
On slave units in series connection, the output voltages of the slave units
alone are displayed (when output is on).
No panel operations are enabled on the slave units. Besides, the Off
Timer, sequence and external control functions of the slave units are
initialized. On slave units in parallel connection, ---- is displayed and
the monitor value is not displayed.
(The monitor value is displayed on the master unit.)
See Sections 6-2. One-Control Parallel Operation and 6-3.
One-Control Series Operation below.
Display on slave unit
in parallel connection
Display on slave unit
in series connection