Thermo Scientific HyPerforma G3Pro Universal Bioreactor Controller Users Guide | 16
Chapter 1 | G3Pro overview
Thermal Control Unit (TCU)8-pin connector that provides
control of the TCU, if used. Used only on vessels larger than 250 L.
TruLogic (required for DeltaV platform)
AUX-1 through 4—5-pin (010 V) auxiliary connectors for control
loops. Note: Spare connectors are available as needed.
AI-1 through 4—5-pin analog (0–20 mA) inputs. They are input
only, with no control capacity.
Vessel Adapter Box (VAB)Enables you to easily switch vessels
according to the requirements of the process.
Auxiliary input pin-out
The connectors marked AUX 1 through 4 are for control loops requiring
feedback from the controller and contain analog input (AI) and analog
output (AO).
Figure 1.5. 5-pin connector.
4
3
2
5
1
The connectors are 5-pin and are set up as follows:
1. Setpoint +
2. Setpoint Return
3. Process Value +
4. Process Value Return
5. Shield
The connectors marked AI 1 through 8 are input-only and are used for
auxiliaries such as CO
²
sensors.
1. +24 V
2. 24 V Return
3. Process Value +
4. Process Value Return
5. Shield