Thermo Electron Corporation VIII-1Orion Star Series Meter
Chapter VIII
Dissolved Oxygen Technique
Dissolved Oxygen Setup Options
The following table describes Dissolved Oxygen setup.
Description Display Range of Values Default
Value
Method
Specific
Dissolved Oxygen
%Saturation Resolution
DO
rES
1
1, 0.1
DO % Saturation
Resolution
0.1
Yes
Dissolved Oxygen
Concentration Resolution
DO
rES
0.1
0.1, 0.01
DO Concentration
“mg/L” resolution
0.01
Yes
Dissolved Oxygen
Barometric Pressure
Compensation Selection
DO
bAr
AUtO
AUtO, mAn
Selects Internal Barometer
or Manual Pressure
AUtO
Yes
Dissolved Oxygen Manual
Barometric Pressure
Setting
DO
PrES
760.0
450.0 - 850.0
Manual Pressure
Compensation Value
760.0
Yes
Dissolved Oxygen Salinity
Correction Selection
DO
SAL
AUtO
AUtO, mAn
Selects Method of Salinity
Correction; Meters w/Cond
Only
AUtO
Yes
Dissolved Oxygen Manual
Salinity Correction Factor
DO
SALF
0
0 - 45
Manual Salinity Correction
Factor
0
Yes
Dissolved Oxygen
Calibration Type Selection
DO
CALt
Air
Air, H
2
O, mAn, SEt0
Air = Water Sat Air,
H
2
O = Air Sat Water,
mAn = Manual,
SEt0 = Zero Cal
Air