Thermo Electron CorporationXIII-8 Orion Star Series Meter
Care and Maintenance
Self-Test Instructions
1. Power up the meter and wait until the SW revision is displayed, then press .
2. The display will then fully light and remain lit.
3. Once you have visual checked that all display segment have lit press again
and this will turn off the display.
4. After visually confirming that all the display elements are turned off, press
again the display will show KEY.
5. Press each key on the meter. If each key isn’t pressed in 5 seconds it will
display “Err” on the middle line & “033” on the bottom line indicating a key
failure. Press to clear the error and complete the self-test.
6. If all the keys are functioning then the meter will restart and go to measurement
mode.
7. If you get the error 034 during the self-test, check that the shorting cap is
installed on the meter.
8. Reject the meter if any errors are reported during self-test.