Controls, Indicators, and Displays
3-17
THE SET SPEED EXCEEDS ROTOR MAX
- ENTER A LOWER SET SPEED
(Future use, if rotors are added.) START was pressed, but the
centrifuge detected a set SPEED value that was too high for
the SET rotor. The centrifuge will not start. This can occur if
you change the SET rotor, but the SET speed is too high for
the newly specified rotor, or if you recall a program that was
saved based on the use of a different, higher-speed rotor.
Make sure the SET rotor specified is the installed rotor, then
enter an rpm or RCF value that does not exceed the maxi-
mum allowable for the rotor, or install and specify a different
rotor, and press START.
THE SET SPEED EXCEEDS ROTOR MAX
- RUN TERMINATED
(Future use, if rotors are added.) This can occur when a QC
RUN or STEP RUN is in progress, and a program is encoun-
tered that has a higher set speed than the maximum speed of
the SET rotor (indicating that the program was saved based
on the use of a different, higher-speed rotor). The SET rotor
cannot be used with the programs as saved. Make sure the
SET rotor is installed, change to an acceptable rotor or
review and change the programs saved before reattempting.
(WatchLog Network™ record: SET SPEED.)
THE SET TIME = 0:00 - ENTER A VALID TIME
START was pressed with a set run duration value of zero.
The centrifuge will not start. Press one of the three run
duration control keys (TIME/HOLD/ACE), input a value,
then press START.