RC12
BP
Operation
4-5
NOTE To prevent accidental erasure of established programs,
if the entered program number has been previously
assigned, the following message will appear:
IN USE - OVERWRITE?
(1=YES, O or ENTER=NO)
To review stored programs before overwriting, refer to
your Program Log Pad or press RECALL. If you press
RECALL, you will have to go back and reestablish the
run parameters before you can store them to memory.
B. Performing a Programmed Run
Read the Safety Information Page at the front of this manual.
1. Press RECALL – the HOME screen will change to display a
program number flashing in the PROG # field.
2. Input the desired program number, then press ENTER – the SET
display and options indicators change to reflect the run param-
eters that were saved to that program number.
3. Prepare the rotor according to the rotor manual instructions.
4. Open the chamber door, wipe the chamber and drive spindle dry
(if moisture is present), and install the rotor, making sure that the
rotor is fully seated and locked to the drive spindle. Install the
rotor cover, if any, then close the chamber door.
NOTE Rotor names are not saved or recalled (in future use, if
new rotors have been added) be sure that the installed
rotor is correct for the recalled program; make sure that
the SET rotor identifies the installed rotor.
5. If using WatchLog Network™ run logging software, be sure it is
running, then press START (START indicator will blink, then
stay lit once the rotor starts to spin).
The rotor will accelerate according to the recalled parameters, and
continue until run termination. At termination, the STOP indicator
will light and the rotor will decelerate to a stop. After it stops, START
and STOP indicators will turn off, and "End" will appear in the RUN
TEMP display and remain there until the door is opened.
NOTE Do not attempt to open the door unless End is in the
SPEED display. Doing so when the interlock is to be
moving could cause the interlock to bind, requiring cy-
cling the main POWER switch several times.