MAINTENANCE
INSPECTING THE OPTOMETRY UNIT MOUNTING SECTION
If the mounting section for the CV-5000 and the optometry unit, which are in use, is loosened,
an unexpected accident may be caused.
To use the instrument safely, it is recommended to inspect it every day.
Point for inspection
Inspect the connecting section between the CV-5000 and the arm.
Raise the CV-5000 instrument as shown below. If an abnormal noise is heard or any part is
unsteady between the arm and the arm mounting shaft, immediately contact the TOPCON
offices listed on the back cover and ask them to repair your instrument.
Raise the instrument vertically. If you raise it slantwise, it is difficult to check if any part is
unsteady because the arm mounting shaft moves horizontally.
NOTICE
Hold the PD unit when moving the head.