EasyManua.ls Logo

TR 900 - Routine Maintenance: Before Use and Daily; Daily Visual Inspection Checklist; Daily Tub Cleaning Procedure; Daily Post-Use Procedure

Default Icon
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hygiene Equipment
Designed for People
40
Routine maintenance
Before every use
Visual inspection:
- Make sure that the bathtub stands securely on the floor.
- Make sure that the tub is in its lowest position.
- Make sure that the tub is in level when raising and lowering the tub.
- Make sure there are no nicks or cracks on the tub which may scratch the patient.
- Make sure that all electrical cords are in good condition and without damages.
- Make sure that the connecting- and shower hoses do not leak.
- Check the water temperature by hand or with a bath thermometer.
Daily maintenance
- Clean the tub after every use.
- Avoid abrasive cleaners which may scratch the surface of the tub and side covers.
- Lower the tub to its lowest position after every use and make sure all valves are closed.
- Clean the strainers on the shower handle by rubbing it with your thumb.
- After use make sure the bathtub is in its lowest position.
- After use, make sure all valves are properly closed.
If your TR 900 is equipped with a cleaning system:
- Make sure there is cleaning liquid in the cleaning container.
- Clean the strainers on the cleaning shower handle by rubbing it with your thumb.
Warning
Do not use Iodine, Bromine and Methylene blue when cleaning the glass fiber tub, this
might discolor the bathtub.
Warning
Do not use Chlorine tablets when cleaning the Stainless-steel tub (-SS), this might cause
rust stains on the tub.