B-6 Model 3775 Condensation Particle Counter
Data Type 2, 3771/72/75:
UX,C,C,C,C,C,C,C,C,C,C,R,R,R,R,R,R,R,R,R,R,F,DTC,T,T,T,T,T,T,T,T,T,T
UX => elapsed time(sec), integer
C => tenth sec concentration, float
R => tenth sec raw counts, integer
F => flowrate(cc/sec), float
DTC => deadtime correction, float
T => tenth sec deadtime(sec), float
SCM Set the operating mode
Params: 0 => Concentration
1 => Totalizer
2 => SMPS
Example: SCM,0 (sets operating mode to concentration)
STS Set saturator temperature
Params: c => 0.0–50.0
Example: STS,39.0 (changes the saturator set point to
39.0 degrees C)
STC Set condenser temperature
Params: c => 0.0–50.0
Example: STC,14.0 (changes the condenser set point to
14.0 degrees C)
STO Set optics temperature
Params: c => 0.0–50.0
Example: STO,40.0 (changes the optics set point to
40.0 degrees C)
SAWR Set the auto water removal function on/off
Params: 0–Off
1–On
Example: SAWR,1 (turns on water removal)
SVO Set analog output voltage
Params: v => 0.000–10.000
Example: SVO,4.482 (sets the output voltage at 4.482
volts)