EasyManua.ls Logo

TSI Instruments 3775 - Page 119

TSI Instruments 3775
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Firmware Commands B-7
SAO Set analog output voltage proportional to concentration.
The analog output is 0 to 10V.
Params: 0 => Off
1 => 1E1
2 => 1E2
3 => 1E3
4 => 1E4
5 => 1E5 (CPC100 only)
Example: SAO,4 (A concentration reading of 1E4 will
equal 10V)
SCOM Setup auxiliary comport
Params: Port => 1,2,3
Baud =>
2400,4800,9600,14400,19200,28800,38400,
57600,115200
Bits => 5,6,7,8
Parity => E, O, N
Stop => 1, 1.5, 2
Example: SCOM,2,9600,7,E,1 (Set 2nd serial port to
9600, 7 bits, Even Parity, 1 Stop bit
SHOUR Set the Real Time Clock Hours (24 hour mode)
Params: hour => 0–23
Example: SHOUR,13 (sets the hour to 13)
SMINUTE Set the Real Time Clock Minutes
Params: min => 0–59
Example: SMINUTE,45 (sets minutes to 45)
SSECOND Set the Real Time Clock Seconds
Params: sec => 0–59
Example: SSECOND,0 (sets seconds to zero)
SYEAR Set the Real Time Clock Year
Params: year => 0–99
Example: SYEAR,6 (sets the year to 2006)
SDAY Set the Real Time Clock Day of the Month
Params: day => 1–31
Example: SDAY,23 (sets the day to the 23rd of the
month)

Table of Contents

Related product manuals