Firmware Commands B-7
SAO Set analog output voltage proportional to concentration.
The analog output is 0 to 10V.
Params: 0 => Off
1 => 1E1
2 => 1E2
3 => 1E3
4 => 1E4
5 => 1E5 (CPC100 only)
Example: SAO,4 (A concentration reading of 1E4 will
equal 10V)
SCOM Setup auxiliary comport
Params: Port => 1,2,3
Baud =>
2400,4800,9600,14400,19200,28800,38400,
57600,115200
Bits => 5,6,7,8
Parity => E, O, N
Stop => 1, 1.5, 2
Example: SCOM,2,9600,7,E,1 (Set 2nd serial port to
9600, 7 bits, Even Parity, 1 Stop bit
SHOUR Set the Real Time Clock Hours (24 hour mode)
Params: hour => 0–23
Example: SHOUR,13 (sets the hour to 13)
SMINUTE Set the Real Time Clock Minutes
Params: min => 0–59
Example: SMINUTE,45 (sets minutes to 45)
SSECOND Set the Real Time Clock Seconds
Params: sec => 0–59
Example: SSECOND,0 (sets seconds to zero)
SYEAR Set the Real Time Clock Year
Params: year => 0–99
Example: SYEAR,6 (sets the year to 2006)
SDAY Set the Real Time Clock Day of the Month
Params: day => 1–31
Example: SDAY,23 (sets the day to the 23rd of the
month)