Counts displayed on the data tab concentration may have slight
rounding errors when comparing all channels to values with selected
channels enabled. The method for calculating concentration is to sum
the raw counts for each location, calculate concentration from sample
volume and then round. This may result in slight rounding errors when
comparing counts with all channels enabled versus concentrations with
selected channels enabled. The methodology utilized is covered in
ISO 14644-1 Annex D.
Button used to change between counts and concentration
displays:
# = number
ft
3
= particles per cubic foot
m
3
= particles per cubic meter
µg = micrograms per cubic meter (mass
concentration)
Export the data to a flash drive. See Export Data Screen
below.
Print data to the optional printer. See Print Data below
Zone where the data was collected.
Location where the data was collected.
Duration of the sampling period.
Date on which the data was collected.
Time at which data was collected.
Temperature at the end of the time the data was collected
(if probe connected during sampling).
Humidity level at the end of the time the data was collected
(if probe connected during sampling).
Alarm threshold was triggered (Yes) or not (No).
Status of the instrument hardware. OK if no issues or error
abbreviation if instrument has a possible issue.
Volume of air that was sampled.