Notes on the Product
Legal requirements
6
Hans Turck GmbH & Co. KG | T +49 208 4952-0 | F +49 208 4952-264 | more@turck.com | www.turck.com
2 Notes on the Product
2.1 Product identification
X
Special version
Special version
X Peak pressure aperture
PS 310 – 1V – 03 – LI2UPN 8 – H1 1 4 1
PS 310
Measuring cell –
Measuring cell
310 Ceramic measuring cell
510 Metal measuring cell
Functional principle
PS Pressure sensor
1V
Measuring range –
Measuring range
Ceramic measuring cell
OV -1 … 0 bar
1V -1 … 1 bar
1 0 … 1 bar
1A 0 … 1 bar absolute
2.5V -1 … 2.5 bar
2.5 0 … 2.5 bar
2.5A 0 … 2.5 bar absolute
Metal measuring cell
10V -1 … 10 bar
10 0 … 10 bar
10A 0 … 10 bar absolute
16 V -1 … 16 bar
16A 0 … 16 bar absolute
25V -1 … 25 bar
25A 0 … 25 bar absolute
40V -1 … 40 bar
100 0 … 100 bar
250 0 … 250 bar
400 0 … 400 bar
600 0 … 600 bar
03
Mechanical version –
Mechanical version
01 G1/4" female thread
02 1/4"-18NPT female thread
03 1/4"-18NPT male thread
04 G1/4" male thread
05 7/16" UNF male thread
08 G1/2" manometer
LI2UPN 8
Electrical version /
Operating voltage
8 18...30 VPC
Output function
2UPN 2 switching outputs/
IO-Link
LI2UPN Current and
switching output/
IO-Link
H1 1 4 1
Electrical connection:
connector
/
Assignment
1 Standard assignment
Number of contacts
4 4 Contacts
Alignment
1 Straight
Design
H1 Connector
M12 × 1
2.2 Scope of delivery
The scope of delivery includes:
n Pressure sensor
n Quick-Start Guide
2.3 Legal requirements
The device is subject to the following EC directives:
n 2014/30/EU (electromagnetic compatibility)
n 2011/65/EU (RoHS Directive)