Setting
Setting via touchpads
24
Hans Turck GmbH & Co. KG | T +49 208 4952-0 | F +49 208 4952-264 | more@turck.com | www.turck.com
Explanation Options Function
ou2
Analog output auto
4-20 4…20 mA
0-20 0…20 mA
20-4 20…4 mA
20-0 20…0 mA
0-10 0…10 V
0-5 0…5 V
1-6 1…6 V
10-0 10…0 V
5-0 5…0 V
6-1 6…1 V
rtio 0.5…4.5 V
SP2
Switching point 2
ou2: Hno/Hnc
Upper limit value at which output 2 changes
its switching state when the pressure increases
rP2
Reset point 2
ou2: Hno/Hnc
Lower limit value at which output 2 changes
its switching state when the pressure drops
FH2
Upper switch point for window
function
ou2: Fno/Fnc
Upper switch point at which output 2 changes
its switching state
FL2
Lower switch point for window
function
ou2: Fno/Fnc
Lower switch point at which output 2 changes
its switching state
ASP
Start point of the analog signal
ou2: Auto/analogvalues/rtio
Pressure value at which the analog signal has
its start point
AEP
End point of the analog signal
ou2: Auto/analogvalues/rtio
Pressure value at which the analog signal has
its end point
EF
Submenu for additional setting
options
See table “Parameters in the EF submenu”
9.1.5 Parameters in the EF submenu (Extended Functions)
Explanation Options Function
Hi
Maximum value memory The highest pressure is stored and can be dis-
played/deleted here.
Lo
Minimum value memory The lowest pressure is stored and can be dis-
played/deleted here.
CoF
Offset adjustment Severe temperature changes in the environ-
ment of the sensor can cause the shifting of
the zero point. In this case, the displayed
measured value will not be zero when the
sensor is in a depressurized state. An offset
value can be set in order to correct this drift:
-5 …+5 % of the measuring range.
dSP1
Switch delay of SP1 0…60 s in increments of 0.1 s (0 = delay time
not active)
drP1
Switch delay of rP1 0…60 s in increments of 0.1 s (0 = delay time
not active)