83
Corrections
6.1 See symptom 20 in para. 10.4.
7.1 make sure water pump is working. Make
sure electrode wires are connected
Possible cause check-up and tests
6.1 Incorrect reading of the reservoir float
switch position.
7.1 Check that the electrodes are sensing
water in the jacket
10. TROUBLESHOOTING (cont.)
10.3 Pre-process malfunction (cont.)
Symptom
6. Pressing the START key
does not start the process
and the error message
ADD WATER is displayed
despite the water reservoir
being full.
7. Unit will not heat but
display is on (except during
vacuum test).