44
9.5 Replacing the Door Gasket
Pull off the gasket from the door groove. Install the new gasket as
described in drawings 1, 2 and 3 above.
Caution!
This gasket
is designed with a trapezoidal cross section. The gasket
should be placed with the widest side towards the door.