(figure 7)
Resistance Measurement (see figure 7)
Measurement Operation (5)
The resistance ranges are: 800.0 , 8.000k , 80.00k ,
800.0k , 8.000M and 80.00M . To measure
resistance, connect the Meter as follows:
1. Insert the red test lead into the
V Hz terminal
and the black test lead into the COM terminal.
2. Set the rotary switch to
, resistance
measurement is default or press
to select
measurement mode.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l The test leads can add 0.1 to 0.2 of error to
resistance measurement. To obtain precision
readings in low-resistance measurement, that is the
range of 800.0 , short-circuit the input terminals
beforehand, using the relative value function button
REL
to automatically subtract the value measured
when the testing leads are short-circuited from the
reading.
l For high-resistance measurement (>1M ), it is normal
taking several seconds to obtain a stable reading.
Select
W Test
blackred
Model UT70C: OPERATING MANUAL
20