Accuracy Specifications (4)
H. Frequency & Duty Cycle
Frequency - At mV
Range
Remarks:
l Input sensitivity as follows:
<100kHz:
100mV rms;
100kHz: 500mV rms;
l Maximum input amplitude: 30V rms.
Frequency - At V
& V Range
Range Resolution Accuracy
Overload
Protection
500V rms(0.02%+1)
0.01Hz
0.1Hz
1Hz
10Hz
1kHz
10kHz
100kHz
1MHz
Range Accuracy Overload Protection
1000V(0.1%+3)10Hz~100kHz
Remarks:
l Input amplitude:
500mV rms.
l Maximum input amplitude: 30V rms.
Frequency - At A
& mA Range
Remark:
l Input amplitude:
3000 digits.
Duty Cycle
Range Accuracy Overload Protection
(0.1%+3)50Hz ~ 10kHz
Range Resolution Accuracy
Overload
Protection
500V rms(1%+5)0.01%1% ~ 99%
10A,250V fast type
glass fuse, 5x20mm.
1A,250V fast type
glass fuse, 5x20mm.