27
Error 1
Instrument unable to
initialize and/or locate the
lamp (S2100UV+only).
Check power supply to see if proper voltage is
selected.
Open cover of instrument and check the connector J3
between CPU board and Lamp-change-over motor .
Lamp-change-over motor or lamp initialization
switch malfunctioning,Check the connector J8 on
CPU board.
Call authorized service engineer
Error 2
Instrument unable to
initialize and/or locate the
filter
Open cove rof instrument and check the connector J2
between CPU board and Filter-driven motor.
Filter-driven motor malfunctioning ,Check the
connector J4 on CPU board.
Call authorized service engineer
Error 3
Monochromator unable to
locate starting position
Monochromator contact switch or grating-driven
motor (WL motor) is malfunctioning
Open cover of instrument and check the connector
J3(S1205) between CPU board and grating-driven
motor.
Check the connector J5 on CPU board
Call authorized service engineer.
Error 4
Instrument unable to locate
“0” order light
Check sample holder. See if holder is properly
positioned and nothing is blocking light path.
Check to see if light is positioned properly on
entrance slit to monochromator.
Check tungsten-halogen lamp. Replace lamp if
necessary.
Call authorized service engineer.
Error 5
Instrument is unable to set
0A/100%T.Energy
Low!” display.
Check sample compartment and remove cuvette or
any object blocking light path.
Place the 4-cell sample holder in the proper position.
See the lamps is ON or OFF via the small hole in the
rear of the instrumen.if OFF,call authorized service
engineer.
Contact an authorized service engineer.
Error 6
Data overflow
The dark current is too high. Check the dark current;
The blanking reference liquid is too concentrated
The light source is off: lamp is not on; light beam
is blocked, alignment is off.
The secondary filter wheel is jammed or the filter
is damaged.
The standard curve is not established properly.

Questions and Answers: