Syndrome Rated capacity cannot be reached
Check Is the TARE INDICATOR on?
Action: Turn the scale off. Remove all loads
from platter and turn on again.
Check Is there anything obstructing the
platter?
Action Remove all obstacles.
Syndrome Weighing result is not accurate
Check Is the scale placed in a level
condition?
Action Adjust the adjustable feet to a level
condition.
Check Is the scale affected by airflow,
vibration or RFI?
Action: Place the scale away from all
disturbances.
Check Is the scale calibrated correctly?
Action Contact your dealer.
Check Is the correct weight unit employed?
Action Refer to 4.2.1 and 5.3 for weight
unit setting and usage.
Syndrome Scale displays
00000
Check Is the scale overloaded?
Action Remove some loads from platter.
Syndrome Scale displays
00000
Check Is the stainless steel platter cover
placed on the platter?
Action Place the stainless steel platter
cover onto the platter.

Related product manuals