EasyManua.ls Logo

Vaisala HUMICAP HMW92 - Set Output Clipping and Error Limit

Vaisala HUMICAP HMW92
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
48 ___________________________________________________________________ M211399EN-G
Set Output Clipping and Error Limit
Use the AOVER command to define the behavior of the analog outputs
when the measured value is outside the scaled output range.
AOVER [channel clip% valid%]<cr>
where
channel
=
Analog output channel, 1 or 2.
clip%
=
Output margin (%) at which the output is clipped.
Range 0 ... 20, default is 0.
valid%
=
Output margin (%) at which the output of the channel
goes into the error state. Range 0 ...
20, default is 5.
The error state is defined using the
AMODE command,
see section Set Analog Output Mode on page 46.
NOTE
These settings have no effect on the measurements shown on the display.
The display will always show the currently measured values, even
outside the scaled output range, as long as the measurement is still
functioning.
For example, first check the analog output settings using ASEL,
AMODE, and AOVER commands:
>pass 9000
>asel
Aout 1 quantity : RH (0.00 ... 100.00 %)
Aout 2 quantity : T (-5.00 ... 55.00 'C)
>amode
Aout 1 range ( V) : 1.00 ... 5.00 (error: 6.00)
Aout 2 range ( V) : 1.00 ... 5.00 (error: 6.00)
>aover
Aout 1 clipping : 0.00 %
Aout 1 error limit : 5.00 %
Aout 2 clipping : 0.00 %
Aout 2 error limit : 5.00 %
The parameter for channel 2 is T, with standard output range 1 ... 5 V and
scaling -5 ... 55 °C. Error state is 6 V, which is set when the measured
value is 5% outside the scaled output range.
Now give the following AOVER command:
>aover 2 10.0 20.0
Aout 2 clipping : 10.00 %
Aout 2 error limit : 20.00 %

Table of Contents

Related product manuals