Chapter 6 ____________________________________________________________ Troubleshooting
VAISALA ________________________________________________________________________ 77
Reverting to Factory Settings Using
Service Port
Use the FRESTORE command to restore the transmitter to factory
settings.
FRESTORE<cr>
NOTE
If you are restoring the factory configuration on a factory customized
analog output model (HMW90), leave DIP number 8 to the
Custom
position to avoid the custom configuration from being overwritted by the
default settings at transmitter reset. For more in
formation about this, see
section
Changing Between DIP and Custom Configuration on page 23.
NOTE
After using the
FRESTORE command, reset the transmitter using the
RESET
command.
Example:
>pass 9000
>frestore
Restoring HTM10 factory parameters
HTM10 factory parameters restored
Restoring HMW93 factory settings
72/72 parameters restored
OK
>reset
Resetting
HMW93 / 1.00.00.0000 / XM90
>
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
helpdesk@vaisala.com. Provide at least the following supporting
information:
- Name and model of the product in question.
- Serial number of the product.
- Name and location of the installation site.
- Name and contact information of a technically competent person who
can provide further information on the problem.
For Vaisala Service Center contact information, see
www.vaisala.com/servicecenters.

Table of Contents

Related product manuals