Chapter 6 ____________________________________________________________ Troubleshooting
VAISALA ________________________________________________________________________ 73
Viewing Error Messages on Serial Line
View Currently Active Errors
Use the ERRS command to view currently active errors on the serial
line:
ERRS<cr>
Example:
>errs
NO ERRORS
View Error Table
Use the ERRT command to view the table of possible transmitter errors.
The table includes error ID, error count since last reset, level, current
state, and error text.
Critical errors require a transmitter reset to recover. Other errors may be
recoverable if their cause is removed.
ERRT<cr>
Example:
>errt
Id: N: Level:State: Error text
1: 0: CRITICAL:OFF: FLASH memory corrupted
2: 0: CRITICAL:OFF: Parameter read (using defaults)
3: 0: CRITICAL:OFF: Parameter write
4: 0: CRITICAL:OFF: HTM10 03 FLASH Corrupted
21: 0: ERROR:OFF: HTM10 04 RH measurement
22: 0: ERROR:OFF: HTM10 05 T measurement
23: 0: ERROR:OFF: HTM10 01 Continuous communication failure
41: 0: WARNING:OFF: HTM10 02 Single Communication failure
42: 0: WARNING:OFF: HTM10 06 Device Descriptor match
43: 0: WARNING:OFF: Factory parameter memory not consistent