EasyManuals Logo

Vatech A9 User Manual

Vatech A9
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
9. Acquiring CBCT Images
vatech A9 (Model name: PHT-30CSS) User Manual 107
English
9.4.3 Patient Positioning - Normal
1. Insert the Normal Bite into the Normal Chinrest and cover it with a Sanitary Vinyl
Cover.
The Sanitary Vinyl Cover is for sole use only. It should be
replaced for each patient. Be sure to use the approved
vinyl cover.
Clean the Chinrest and the Bite with a non-alcohol-based,
non-corrosive cleaning solution and wipe with a dry towel
before the next patient.
2. Guide the patient to a chair (sold separately) in the center of the equipment.
3. Guide the patient to sit in the center of the equipment
If a chair with a backrest is used, make sure that the
patient's back is away from the backrest of the chair for
proper positioning.
Normal Bite
Sanitary Vinyl Cover

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals