2. X-ray Exposure Guide
< Table 4. Detailed Exposure Guidance Table>
Ref.
Exposure condition Dose
Signal Level
60kvp
2mA
60kVp
6mA
65kVp
5mA
Patient
Cone to skin
Section
-
-
No
Filter
-
-
4mm
Al filter
-
-
Adult
8cm
AnyRay
Adult
8cm
VX 70
Adult
8cm
ESX-
Series
Upper
Approximate Exposure Time
Upper
Lower
Jaw/
Anterior
Upper
Jaw/
Canine
Upper
Jaw/
Posterior
Lower
Jaw/
Anterior
Lower
Jaw/
Canine
Lower
Jaw/
Posterior
2800
2800
3200
2400
2400
2800
1800
1800
2100
1400
1400
1800
160 0.20 0.16 0.22
160 0.20 0.16 0.22
200 0.25 0.20 0.27
130 0.15 0.12 0.18
130 0.15 0.12 0.18
160 0.20 0.16 0.22
For larger body types : increase the source current by 25%
For children : reduce the source current by 20%
For edentulous patients : reduce the source current by 20%