Appendix
9. Code 8
Message : Bad Pixel's map correction error.
Solution : Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service.
10. Code 9
Message: Bad Pixel's map correction error.
Solution: Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service.
11. Code 10
Message :. Can't find dark frame.
Solution : Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service
12. Code 11
Message: Can't find bright frames for calibration
Solution: User canceled image capture. Please try again.
13. Code 12
Message: Canceled image capturing.
Solution : User canceled image capture. Please try again.
14. Code 13
Message: Wrong image processing parameters.
Solution: Check X-ray source.
15. Code 14
Message: Error in configuration file.
Solution: Please re-instal acquisition software.
16. Code 15
Message: Image process library was damaged.
Solution: Please re-instal acquisition software.