8 Acquiring CT Images
F. Gender selection will be made automatically according to the patient
information registered in EasyDent / EzDent-i. Nevertheless, ensure that this
information is correct.
Age Group VATECH’s Standard
Child
12
Adult
Man
13
Woman
G. Select X-ray intensity.
NOTE
Depending on the circumference of the patient’s head,
X-ray intensity may be classied as Hard, Normal, or Soft :
Soft ≤ Normal ≤ Hard
Age
Group
Average head
circumference
Range
(cm)
classication of head
circumference
Child
53
±
3 cm
> 53
±
3 Hard
53
±
3 Normal
53
±
3 < Soft
Adult
56
±
3 cm
> 56
±
3 Hard
56
±
3 Normal
56
±
3 < Soft
H. A default tube voltage (kVp) and current (mA) value will be displayed based
on the patient’s gender and X-ray intensity. If necessary, you can perform
manual adjustments by clicking the arrows to the right of each number.
[PI3DG_130U_44A_en]User Guide.indd 114 2016-05-24 오후 4:19:10