EasyManua.ls Logo

Vatech PHT-60CFO - Page 158

Vatech PHT-60CFO
167 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PaX-i3D Green (PHT-60CFO) User Manual
158
Appendices
Guidance and manufacturer’s declaration - electromagnetic immunity
The PaX-i3D Green (Model: PHT-60CFO) is intended for use in the
electromagnetic environment specied below.
The customer or the user of the PaX-i3D Green (Model: PHT-60CFO)
should assure that it is used in such an environment.
Immunity
Test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic environment
-guidance
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80MHz V1=3Vrms
Portable and mobile RF
communications equipment
should be used no closer to
any part of the PaX-i3D Green
(Model: PHT-60CFO), including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
d P
= [ ]
3,5
V
1
d P
= [ ]
3,5
E
1
80 MHz to 800MHz
d P
= [ ]
7
E
1
800 MHz to 2,5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of the equipment marked
with the following symbol :
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5GHz
E1=3V/m
[PI3DG_130U_44A_en]User Guide.indd 158 2016-05-24 오후 4:19:24