6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Select " Show table" to display and edit the linearisation points.
Select " Linearisation - New"toentertherstpoint.
Select " Determine count rate"toentertherstpoint.
The determination of the actual count rate lasts 2 minutes. After the
countratehasbeendetermined,youcanacceptthevalue(ct/s).
Thecountrateisstatedinct/s.Thisisthenumberofcountspersec-
ond, i.e. the measured radioactive radiation dose actually reaching
the sensor.
Enternowthecorrespondingdensityvalue(g/cm
3
).
You thus assign a corresponding density to the actual count rate.
Note:
If possible, take a product sample and determine the density simulta-
neously in your sampling position.
Ithasprovenmarkingtheproductsmapleswithdateandrespec-
tivecountrate.Hencethevaluescanbeassignedlateronwithout
problems.
Acceptthevaluepairwith"OK".
Enterasmanylinearizationpointsaspossible.Youcaninuence
the accuracy of the density measurement by doing this. The more
linearizationpointsyouenterandthehigherthedierencebetween
the density values, the more reliable the measurement.

Table of Contents

Other manuals for Vega MINITRAC 31

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vega MINITRAC 31 and is the answer not in the manual?

Vega MINITRAC 31 Specifications

General IconGeneral
BrandVega
ModelMINITRAC 31
CategoryAccessories
LanguageEnglish

Related product manuals