EasyManua.ls Logo

Vega MINITRAC 31 - 6.7 Parameter adjustment - Real value correction

Vega MINITRAC 31
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
In the menu item " Lock/unlock adjustment", you can protect the sen-
sorparametersagainstunauthorizedorinadvertentmodication.The
sensorislocked/unlockedpermanently.
Withlockedinstrument,onlythefollowingadjustmentfunctionsare
possiblewithoutenteringaPIN:
Selectmenuitemsandshowdata
Read data from the sensor into the display and adjustment module
Before you lock the sensor in unlocked condition, you can modify the
four-digit PIN number.
Keep the entered PIN number in mind. Operation of the sensor is only
possiblewiththisPINnumber.
Caution:
WhenthePINisactive,adjustmentviaPACTware/DTMaswellas
other systems is also blocked.
In delivery status, the PIN is 0000.
Callourservicedepartmentifyouhavemodiedandforgottenthe
PIN.
6.7 Parameter adjustment - Real value correction
To correct a level measurement, a real value correction can be carried
outwhenacertainlevelisreached.Thisisforexamplenecessaryin
caseofrecurringbuilduponthevesselwall.
The level measurement can thus be reset to the correct real value.
The level measuring instrument acts as Primary and the MINITRAC
31 operates as Secondary.
Firstofall,denethefunctionoftheSecondaryinstrumentsbefore
youdenethePrimaryinstrument.ThePrimaryinstrumentcanthus
immediatelyrecognizetheconnectedSecondaries.
Forthispurpose,theSecondaryinstrumentmustbedenedas"Real
value correction". Select under the menu item " Setup - Application"
the function "Real value correction".
Theaddresssetting(MGC)ontheSecondaryinstrumentscanbe
freely selected. Only the address "0 - 0" is reserved for the Primary
instrument.
The Primary instrument must have the function "Level". For this pur-
pose, select under the menu item " Setup - Application" the function
"Level".
Settheaddresssetting(MGC)onthePrimaryinstrumentto"0-0".
Lock/Unlock adjustment
Real value correction

Table of Contents

Other manuals for Vega MINITRAC 31

Related product manuals