59
6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments:Instrumentunit,reset,date/time,copying
function
Info:Instrumentname,hardwareandsoftwareversion,dateofmanu-
facture, instrument features
Check if the correct language is already set for the display. If not,
you can change the language in the menu item " Display - Menu
language".
StartwiththesetupofMINITRAC31.
In the main menu item " Setup", the individual submenu items should
beselectedoneaftertheotherandprovidedwiththecorrectparam-
eters to ensure optimum setting of the measurement. The procedure
isdescribedinthefollowing.
Stickwiththenormalsequenceofthemenuitemsascloselyaspos-
sible.
6.6.1 Setup
In this menu item you can adjust the MINITRAC 31 to the isotope
installed in the source container.
Forthispurpose,checkwhichisotopeisinthesourcecontainer.You
canndthisinformationonthetypelabelofthesourcecontainer.
With this selection, the sensitivity of the sensor is adapted in an opti-
mumwaytotheisotope.
The MINITRAC 31 requires this information for the decay compensa-
tion. An annual recalibration is hence not necessary.
Enter the requested parameters via the appropriate keys, save your
settingswith[OK]andjumptothenextmenuitemwiththe[ESC] and
the [->] key.
Enter here, the respective application.
This menu item enables adaptation of the sensor to the requested ap-
plication.Youcanchoosebetweenthefollowingapplications:"Level",
" Density", " Limit level", " X-ray alarm" or " Real value correction".
Procedure
Isotope
Application