57
6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Menu Menu item Default value
Setup Measurement loop name Sensor
Isotope Cs-137
Application Limit level
Adjustment mode Single point adjustment
Adjustment - uncovered 90000ct/s
Adjustment - covered 9000ct/s
onlywithtwo-pointadjustment
Delta I 90 %
Background radiation 0ct/s
Temperature unit ° C
Adjustment mode 1-point
Adjustment uncovered 900000ct/s
Delta I 90 %
Mode - Relay Overllprotection
Lock adjustment Released
Display Language Selected language
Displayed value Pulse rate
Display unit ct/s
With this function
•
Load parameter adjustment data from the sensor into the display
and adjustment module
•
Write parameter adjustment data from the display and adjustment
module into the sensor
The copied data are permanently saved in an EEPROM memory in
the display and adjustment module and remain there even in case of
apowerfailure.Fromthere,theycanbewrittenintooneormoresen-
sors or kept as backup for a possible sensor exchange.
Note:
Before the data are copied into the sensor, a check is carried out to
determineifthedatatthesensor.Ifthedatadonott,afaultsignal
istriggered.Whendataarebeingwrittenintothesensor,thedisplay
showswhichinstrumenttypethedataoriginatefromandwhichTAG
number this sensor had.
6.5.5 Info
Inthismenuyouwillndthefollowingmenuitems:
•
Instrumentname-showsinstrumentnameandserialnumber
Copy instrument settings
Info