45
14 Supplement
VEGAPULS C 21 • Modbus and Levelmaster protocol
58343-EN-200806
14 Supplement
14.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions which are included in delivery are valid for
approvedinstruments(e.g.withExapproval).Thesedatacandierfromthedatalistedherein,for
example regarding the process conditions or the voltage supply.
All approval documents can be downloaded from our homepage.
Materials and weights
Materials, wetted parts
Ʋ Antenna,processtting PVDF
Ʋ Counter nut
3)
PP
Ʋ Process seal FKM
4)
Materials, non-wetted parts
Ʋ Housing PVDF
Ʋ Cable entry seal NBR
Ʋ Connection cable PUR
Weight
Ʋ Instrument 0.7 kg (1.543 lbs)
Ʋ Connection cable 0.1 kg/m
Processtting Thread G1½, R1½, 1½ NPT
Mounting connection Thread G1, R1, 1 NPT
Torques
Torque counter nut max.
7 Nm (5.163 lbf ft)
Input variable
Measured variable The measured variable is the distance between the
antenna edge of the sensor and the medium surface.
The antenna edge is also the reference plane for the
measurement.
3)
G type threaded connections only
4)
G type threaded connections only