EasyManuals Logo

Vetek JIK-6 Series User Manual

Vetek JIK-6 Series
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Service Manual Indicator JIK-6 Page 18
A/D board defect.
A/D board is prone to interference, esp. A/D IC;
please clean the board by use of stain removal oil. If
the problem still exists, replace A/D board
.
Handling of Special
Cases such as
Machine Down,
Failure to Power
On/Off as Usual
As shown in Fig.1 on the right,
remove the rear screw of the
display head by use of a cross
head screwdriver.
As shown in Fig.2 on the right,
gently press RESET key inside
the display head by use of a bar
strip and the scale will be
restarted automatically.
How to “Clear
Memory”
Press
key to power on, the display window will show id, then press ,
, and in turn, finally, hold down key (for about 2 seconds),
the window will show
, press key to restart the scale and the
memory will be cleared. Note: After clearing memory, you must at least perform
model setting and linear calibration; otherwise, you cannot use the scale.