8. SOFTWARE
8.1. Overview of the Menus
The Mistral Evolution has the following menus:
PARAMETER Used to set the different parameters
CONFIGURE Only available to trained individuals; password protected
DIAGNOSTIC Used to see the different values of the sensors
ALARM HISTORY Shows the history of alarms
SET CLOCK Sets the clock
SPECIAL FUNCTIONS Enter a password to access specic menus
8.2. Parameters
ITEM DESCRIPTION OPTIONS/RANGE DEFAULT
Pressure Unit Sets desired unit of measure for pressure bar/kPa/psi bar
Temperature Unit Sets desired unit of measure for temperature ºC/ºF ºC
Flow Unit Sets desired unit of measure for ow rate L/min - ft3/min L/min
Auto Start Choose to automatically start production upon
powering on or not
Yes/No No
Display Contrast Sets desired contrast of display 1-10 5
Beeper Sets alarm beep on/off Yes/No Yes
Pre-Alarms in List Choose to see pre-alarms in alarm history or not Yes/No No
External Tank Activated only for external tank. Using this mode,
the generator will stop when tank reaches “P.
Max. Tank” pressure and will start when pressure
drops below minimum level of “P. Min. Tank”
Yes/No No
P. Min. Tank Sets minimum pressure reached in external tank
to restart the generator
1-8 1 bar
P. Max. Tank Sets maximum pressure reached in external tank
to stop the generator
6-9 8 bar
Standby Purge Restart every 24 hours if the generator is in
standby, in external tank mode
Yes/No No