14
MISTRAL EVOLUTION
USER MANUAL
www.v icidbs.com
Version 1.03.0000
8.3. Diagnostic
ITEM DESCRIPTION
Temperature Shows the instant temperature in the cabinet
Outlet Pressure Shows the instant outlet pressure
Int. Tank Pressure Shows the instant pressure in the internal reservoir
Comp. Work Time Shows the work time of the compressor
Service 4000h Shows the work time for the 4000h service
9. OPTIONAL EXTERNAL TANK
In some situation, the LCMS devices do not need the full ow of Nitrogen. This is also the case if the LCMS
stops. In such situation it does not make sense to have the Mistral Evolution run continuously and throw away
the produced Nitrogen.
The Mistral Evolution is able to handle an external tank management. The Mistral Evolution will run until the
high set pressure (P Max Tank) is reached. Then the Mistral Evolution will stop until the low set pressure (P Min
Tank) is reached.
9.1. Size of Tank
The size of the tank is very critical. If at the same time...
the tank is to small
the pressure difference = (P Max Tank)- (P Min Tank) is small
the requested ow is high
...then the Mistral Evolution will start and stop too often and this will damage the compressors of the device.
That’s the reason, once the Mistral Evolution has stopped, he will not restart before 300s (5 minutes).