Vitalograph Spirotrac Instructions for Use
IFU Part No.: 07402
Issue 22
_______________________________________________________________________________________________________________
Page 65 of 78 DT_0006-15
Possible Causes:
(In probable order)
• Ensure tubing is connected correctly. Ribbed side
of the tubing should be connected to the ribbed half
of the connector on the Vitalograph device.
Problem Fault Symptoms: • Incorrect or no volume measurements
Possible Causes:
(In probable order)
• Ensure tubing is connected correctly. Ribbed side
of the tubing should be connected to the ribbed half
of the connector on the Vitalograph device.
• Ensure that the connectors are clear of obstruction
or dirt and that they are inserted fully.
• Ensure the tubing is not kinked or squeezed.
Problem Fault Symptoms: • Excessive calibration drift
Possible Causes:
(In probable order)
• Clean the flow head thoroughly.
• Contact the nearest dealer for replacement.
Problem Fault Symptoms: • Test performed but does not show on screen
Possible Causes:
(In probable order)
• Ensure correct device is selected in Tools >Device.
• Ensure device is connected to PC correctly.
• Ensure tubing is connected between flow-head and
device correctly (same colour connector at both
ends).
Problem Fault Symptoms: • Report does not print all tests
• Report does not print some parameters
Possible Causes:
(In probable order)
• Ensure the correct report settings are set-up in
Tools > Options > Reports.
• Ensure the parameters you require are selected
from Tools > Options > Parameters.