4.Advanced User Guidebook
A→B ,+PulseCount, -PulseCount, RiseEdgeCnt, FallEdgeCnt, Area, and
Cycle Area.
The "Automatic Measurements" menu is described as the following table:
Add the selected measure types (shown at
the left bottom, you could only add 8 types at
most)
Hide the window of measures
Show all the measures of CH1 on the screen
Show all the measures of CH2 on the screen
Show all the measures of CH3 on the screen
Show all the measures of CH4 on the screen
Press to show the left menu, turn the M knob
to select the type need to be deleted, press
Remove again to remove the selected
measure type.
Measure
Only if the waveform channel is in the ON state, the measurement can be
performed. The automatic measurement can not be performed in the following
situation: 1) On the saved waveform. 2) On the Dual Wfm Math waveform. 3) On
the Video trigger mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1. Push the Measure button to show the right menu.
2. Select CH1 in the right menu.
3. In the left Type menu, turn the M knob to select Period.
4. In the right menu, select Add. The period type is added.
5. In the left Type menu, turn the M knob to select Frequency.
6. In the right menu, select Add. The frequency type is added.
The measured value will be displayed at the bottom left of the screen
automatically (see Figure 4- 12).