Current:
Measurement range .................................0.02 - 19.00 A
Tolerance .................................................. < ±2 % @ >0.5 A
< ±5 % @ 0.05 - 0.5 A
< ±0.005 A @ <0.05 A
Resolution .................................................0.001 A @ 0 - 9.999 A
0.01 A @ 10.00 - 19.00 A
Power:
Measurement range .................................0.2 - 3999 W
Tolerance .................................................. < ±2 % @ >10 W
< ±10 % @ 3 - 10 W
< ±0.3 W @ <3 W
Resolution .................................................0.001 W @ 0.3 - 9.999 W
0.01 W @ 10.00 - 99.99 W
0.1 W @ 100.0 - 999.9 W
1 W @ 1000 - 3999 W
Power factor:
Measurement range .................................0.000 - 1.000
Tolerance .................................................. < ±5 % @ >0.5
< ±25 % @ <0.5
Resolution .................................................0.001
67