| Voltage Accuracy | ±0.1% of reading |
|---|---|
| Current Accuracy | ±0.1% of reading |
| Power Accuracy | ±0.2% of reading |
| Accuracy | ±0.1% of reading |
| Harmonic Analysis | Up to 50th harmonic |
| Operating Temperature | 0°C to 50°C |
| Power Supply | 100-240 V AC, 50/60 Hz |
| Current Range | 20 A |
| Display | LCD |
| Interfaces | GPIB |
| Weight | 10kg |
Highlights the danger of electric shock and advises user responsibility for installation compliance and qualified personnel.
Instructs users to check the line input voltage setting before connecting the analyzer to the supply.
Lists essential safety measures, including earthing, power connection order, servicing by qualified personnel, and lead condition.
Illustrates various connection diagrams for single-phase and three-phase measurements, including wiring modes.
Explains how to enter and navigate the analyzer's menu system, select menus, and change options.
Details the calibration process, including password entry and setting calibration values for voltage and current.
Guides through the calibration method, including password entry, setting voltages, and entering readings.
Details the equipment and connections required for voltage calibration of the analyzer.
Describes the procedure for current calibration using internal shunts.
Details the specific setup for calibrating current using an external shunt connection.
Outlines the steps for calibration, including password entry, voltage/current setup, and busy status.
Details the process for calibrating the current measurement ranges (AR1 to AR8).
Details voltage ranges, accuracy, crest factor, input impedance, and common mode voltage effects.
Provides specifications for current ranges, accuracy (excluding neutral), crest factor, and input impedance.
Details display range and accuracy for Power, Apparent Power, and Reactive Power measurements.
Covers specifications for Power Factor range/accuracy and Crest Factor range/accuracy.
Provides specifications for Harmonics (bandwidth, accuracy) and Total Harmonic Distortion (THD) range/accuracy.
Declaration by Voltech Instruments Ltd for the PM100, confirming compliance with Safety and EMC directives.
Details precautions for conducted/radiated emissions and immunity, including ferrite use and recovery from disruptions.