15 VWR Mega Star 600 / 600R ver b, rel 17.07.2020
VWR Mega Star 600R
Environmental Conditions -Use in interior spaces
-Altitudes of up to 3000 m above sea level
-Max. relative humidity 80 % up to 31 °C;
decreasing linearly to 50 % relative humidity at 40 °C
Environmental Conditions during Storage
and Shipping
Temperature: -10 °C to +50 °C
Humidity: 15 % to 85 %
Permissible Ambient Temperature during
Operation
+2 °C to +35 °C
Heat Dissipation
1
230 V, 50 / 60 Hz
0.35 kWh/h; 1170 Btu/h; 1140 kJ/h
Overvoltage Category II
Pollution Degree 2
IP 20
Running Time 99 h 59 min 50 sec, hold
Maximum Speed n
max
17850 rpm
Minimum Speed n
min
300 rpm
Maximum RCF Value at n
max
30279 x g
Noise Level at Maximum Speed
1
< 52 dB (A) (measured with a MicroClick 24 x 2 rotor),
< 56 dB (A) (measured with a TX-150 rotor)
Maximum Kinetic Energy 10.1 kJ
Temperature Setting Range -10 °C to +40 °C
Dimensions
Height (open door / closed door)
Width
Depth
700 mm / 320 mm
460 mm
670 mm
Weight
2
74 kg
1
Front Side Measurement, 1 m in front of the instrument at 1.6 m height
2
Without Rotor
Refrigerant
Order No. Centrifuge Refrigerant Quantity Pressure GWP CO
2
e
521-1894 VWR Mega Star 600R R-134a 0.315 kg 21 bar 1430 0.45 t
521-1895 VWR Mega Star 600R Micro R-134a 0.315 kg 21 bar 1430 0.45 t
Contains fluorinated greenhouse gases in a hermetically sealed system.