EasyManua.ls Logo

VWR Mega Star 600 - Refrigerant

VWR Mega Star 600
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15 VWR Mega Star 600 / 600R ver b, rel 17.07.2020
VWR Mega Star 600R
Environmental Conditions -Use in interior spaces
-Altitudes of up to 3000 m above sea level
-Max. relative humidity 80 % up to 31 °C;
decreasing linearly to 50 % relative humidity at 40 °C
Environmental Conditions during Storage
and Shipping
Temperature: -10 °C to +50 °C
Humidity: 15 % to 85 %
Permissible Ambient Temperature during
Operation
+2 °C to +35 °C
Heat Dissipation
1
230 V, 50 / 60 Hz
0.35 kWh/h; 1170 Btu/h; 1140 kJ/h
Overvoltage Category II
Pollution Degree 2
IP 20
Running Time 99 h 59 min 50 sec, hold
Maximum Speed n
max
17850 rpm
Minimum Speed n
min
300 rpm
Maximum RCF Value at n
max
30279 x g
Noise Level at Maximum Speed
1
< 52 dB (A) (measured with a MicroClick 24 x 2 rotor),
< 56 dB (A) (measured with a TX-150 rotor)
Maximum Kinetic Energy 10.1 kJ
Temperature Setting Range -10 °C to +40 °C
Dimensions
Height (open door / closed door)
Width
Depth
700 mm / 320 mm
460 mm
670 mm
Weight
2
74 kg
1
Front Side Measurement, 1 m in front of the instrument at 1.6 m height
2
Without Rotor
Refrigerant
Order No. Centrifuge Refrigerant Quantity Pressure GWP CO
2
e
521-1894 VWR Mega Star 600R R-134a 0.315 kg 21 bar 1430 0.45 t
521-1895 VWR Mega Star 600R Micro R-134a 0.315 kg 21 bar 1430 0.45 t
Contains fluorinated greenhouse gases in a hermetically sealed system.