32 VWR Mega Star 600 / 600R ver b, rel 17.07.2020
Error number Error message Troubleshooting
E-019 Rotor Unknown Restart the centrifuge.
Check, if the rotor is qualified for the VWR Mega Star 600 / 600R
Centrifuge.
If the error message appears again, inform the service.
E-025;
E-027
Read Manual Check, if the centrifuge door is blocked.
Restart the centrifuge.
If the error message appears again, inform the service.
E-029;
E-045
Read Manual Check, if a rotor is installed.
Check, if the rotor is qualified for the VWR Mega Star 600 / 600R.
Restart the centrifuge.
If the error message appears again, inform the service.
E-030 Power Failure Check the power supply of the centrifuge. Make sure not to operate
too many devices at one power source.
Let the centrifuge cool down for 15 minutes.
If the error message appears again, inform the service.
E-098 Imbalance
Load
Check the load placed in the rotor.
Check that the rotor cross bolts are greased well.
Restart the centrifuge.
If the error message appears again, inform the service.
E-060 Temp Low! CAUTION
Icy metal parts!
Restart the centrifuge.
If the error message appears again, inform the service.
E-046 Door Open! Restart the centrifuge.
If the error message appears again, inform the service.
E-099 Set Speed Too
High
The installed rotor is not rated for the programmed speed.
Check the programmed speed.
When to contact customer service
If you need to contact customer service, please provide the order no. and the serial no. of your centrifuge.
This information can be found on the nameplate at the back near the inlet for the power supply cable.
In addition, the customer service also needs the Software ID and the NVRAM ID. Both are available in the