16 VWR Mega Star 600 / 600R ver b, rel 17.07.2020
Mains supply
The following table contains an overview of the electrical connection data. This data is to be taken into
consideration, when selecting the mains connection socket.
Unit VWR Mega Star 600 VWR Mega Star 600R
Mains Voltage 230 V ±10 % 230 V ±10 %
Frequency 50 / 60 Hz 50 / 60 Hz
Rated Current 2 A 4 A
Power Consumption 310 W 750 W
Equipment Fuse 5 AT 15 AT
Building Fuse 16 AT 16 AT
Directives, Standards and Guidelines
Region Directive Standard
Europe
230 V, 50 / 60 Hz
Machinery Directive 2006/42/EU
Low Voltage Directive 2014/35/EU
EMC Directive 2014/30/EU
RoHS 2011/65/EU
EN 61010-1:2010+A1:2016
EN 61010-2-020:2016
EN 61326-1:2006 (Class B)